The INFINEED lt;precision plus lt; 3-way DEHP free, BPA free PVC extension has been designed to minimise its environmental impact, using only raw materials that are free from endocrine disruptors (elimination of components, including DEHP or BPA). The extension has been tested at pressures well above those recommended for gravity infusion; laboratory tests have shown a pressure resistance of approximately 10 bar.
Extension with:
- BPA-free 3-way stopcock: Lipid-resistant (96 hours to 7 days) with large valve stopcock and arrows indicating the direction of flow, permanently attached to the extension tube.
- Tubing: PVC (DINCH) designed for comfort, optimal flow adjustment consistency and to prevent kinks. Diameter: 2.5 x 4.1mm
- End cap: fixed Luer Lock
- End stopper: 0.8 µm purge filter with hydrophobic membrane
Resistance to pressure of 2 bar according to the claim of compliance with reference standard NF EN ISO 8536-9, 10 bar according to tests performed by Didactic (see scientific and technical file: Lipid-resistant from 96 hours to 7 days)
Made in China.
Contact
with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation
time, and while the luer lock is connected, therefore under physical stress,
presents a significant risk of cracking of the luer lock, and should be avoided
as much as possible.
Indications
This medical device is intended for the parenteral administration of injectable preparations (drugs, nutrients) by gravity, and/or for active infusion devices whose pressure does not exceed 2 bars.
According to good perfusion practice, the infusion line can remain in place for up to 96 hours (CODIMS, OMEDIT CENTRE-VAL-DE-LOIRE, Socle de connaissance de la perfusion - SFAR, EUROPHARMAT, [...]).
In addition, the SF2H reminds us not to change the main infusion line before 4 days, without exceeding 7 days.
This device has been tested for pressure, traction and lipid resistance for up to 7 days.
The length of time the infusion line should be maintained should be discussed and validated in each facility, in order to establish an internal protocol.
Dispose according to regulation in member state.
Used materials
PVC DINCH (DEHP<0.1%(m/>)): Tubing - Polypropylene: Protective cap, stopcock stopper - Polyethylene: Stopcock valve - Tritan® BPA-free: Stopcock body - Acrylonitrile butadiene styrene: Fixed male Luer Lock, fixed female Luer lock
Labeling
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
medical device
sterile
conformity
economic operator