LINEAR EXTENDER PVC WITHOUT DEHP SYRINGE PUSH 1.5X3.0MM MALE FEMALE LUER LOCK FIXED

  • PVC extension 1.5x3.0mm for electric syringe pump
  • Low container/content interaction
  • Good pressure resistance (2 Bar)

Description

Extender with:

High quality tube: flexible anti-kink PVC, DEHP-free
Reliable and secure: Fixed Luer Lock tips, male female tip
Pressure resistance: 2000 hPa (2 Bars , 29 psi)

Internal diameter: 1.5mm

External diameter: 3.00mm

Contact with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation time, and while the luer lock is connected, therefore under physical stress, presents a significant risk of cracking of the luer lock, and should be avoided as much as possible.

In accordance with the recommendations of the SF2H (2019), disinfecting the connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective in disinfecting inert materials and there is much less risk of altering plastic materials.



Technical
information

Indications 
This medical device is intended for the parenteral administration of injectable preparations (medications, nutrients) by electric syringe pump.

Used materials 
Capuchon: PE Luer Lock femelle: PC Tube: PVC Luer Lock male: ABS Capuchon: PE

Labeling 
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable

medical device 

  • MD class : IIa
  • MD notified body : 0197
  • EU directive/regulation : DIRECTIVE 93/42/CEE
  • annexe : V

sterile 

  • sterile : yes
  • sterilization method : ethylene oxide

conformity 

  • Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
  • Infusion equipment for medical use — Part 9: Fluid lines for single use with pressure infusion equipment

Logistics
information

Transport conditions 

Protected from light and humidity

Conditioning Parts quantities Dimensions in cm Weight in kg Ean 13 of conditioning
UNIT 1

Extender with:

High quality tube: flexible anti-kink PVC, DEHP-free
Reliable and secure: Fixed Luer Lock tips, male female tip
Pressure resistance: 2000 hPa (2 Bars , 29 psi)

Internal diameter: 1.5mm

External diameter: 3.00mm

Contact with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation time, and while the luer lock is connected, therefore under physical stress, presents a significant risk of cracking of the luer lock, and should be avoided as much as possible.

In accordance with the recommendations of the SF2H (2019), disinfecting the connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective in disinfecting inert materials and there is much less risk of altering plastic materials.



Indications 
This medical device is intended for the parenteral administration of injectable preparations (medications, nutrients) by electric syringe pump.

Used materials 
Capuchon: PE Luer Lock femelle: PC Tube: PVC Luer Lock male: ABS Capuchon: PE

Labeling 
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable

medical device 

  • MD class : IIa
  • MD notified body : 0197
  • EU directive/regulation : DIRECTIVE 93/42/CEE
  • annexe : V

sterile 

  • sterile : yes
  • sterilization method : ethylene oxide

conformity 

  • Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
  • Infusion equipment for medical use — Part 9: Fluid lines for single use with pressure infusion equipment

Transport conditions 

Protected from light and humidity

Conditioning Parts quantities Dimensions in cm Weight in kg Ean 13 of conditioning
UNIT 1

Associated products

DEHP-FREE PHOTOPROTECTOR SYRINGE PUSH 1.5X3.0MM MALE FEMALE LUER FIXED LOCK

Associated products

DEHP-FREE PHOTOPROTECTOR SYRINGE PUSH 1.5X3.0MM MALE FEMALE LUER FIXED LOCK

Request for technical documentation

  • Didactic formally undertakes not to sell, rent or exchange any personal information. In application of articles 39 and following of the law of January 6, 1978 modified relating to data processing and freedoms (CNIL), you have the right to access, delete and rectify information concerning related to you.