You are:
Rotary flow regulators have a standardized flow stability of +/- 10% over time (ISO8536-14 standard), and are therefore considered lt;precision lt; regulators. Contact us to obtain the good use sheet. Contact us to obtain the correct use sheet.
It should be noted that many factors influence flow regulation (viscosity, height, blood pressure, etc...): drop-counting and regular monitoring are therefore good practice when using these precision regulators.
Extension line with :
- High-quality tubing: flexible, anti-plication PVC
- Precision rotary flow regulator: Smooth, easy adjustment
- Constant, reliable flow thanks to graduated rotary flow regulator.
- Optimum technical features:
Movable Luer Lock tip
Pressure resistance: 50 KPA (0.5 Bars - 7.25psi)
Dead volume: 4.38ml.
Graduation from 0 to 250 mL/h.
Graduations are indicative and reliable up to +/- 20% from 20ml/h upwards.
Thus, counting the drops in drip chamber remains mandatory to ensure that the flow follows the medical prescription.
Made in China.
Contact with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation time, while the luer lock is connected, and therefore under physical stress, presents a significant risk of luer lock cracking, and should be avoided wherever possible.
In line with SF2H (2019) recommendations, disinfecting connectors (when necessary) with 70° ethyl alcohol is always preferable, as it is just as effective at disinfecting inert materials, and with much less risk of altering plastics.
Indications
This medical device is designed for the parenteral administration of injectable preparations (drugs, nutrients).
This device enables precise gravity administration of injectable parenteral preparations: drugs, solutions, parenteral nutrition.
Do not use with high-risk drugs or drugs with narrow therapeutic margins (catecholamines, morphine, heparin, insulin, potassium chloride, anticancer drugs requiring infusion pump administration, and all pharmaceutical specialties whose R.C.P. includes the need for infusion pump or syringe pump administration).
Source : OMEDIT CENTRE -- Les bonnes pratiques de perfusion : Module lt; La précision du débit de perfusion lt;.
Used materials
Capuchon : PE
Luer Lock femelle : PC
Tube : PVC
Régulateur de débit : ABS / TPE
Site d'injection : ABS / Polyisoprène synthétique
Luer Lock mâle : ABS
Capuchon : PE
Labeling
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
medical device
sterile
conformity
Transport conditions
Protected from light and humidity
| Conditioning | Parts quantities | Dimensions in cm | Weight in kg | Ean 13 of conditioning |
| UNIT | 1 | 21 x 13 x 0 | 06972620110510 | |
| BAG | 25 | 55 x 30 x 5 | 0.510 | |
| CASE | 150 | 39 x 42 x 37 | 16972620110517 |